We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

We describe a low field, bench-top MRI system for small samples, based on a permanent magnet. Signals are received at 830 kHz using a tuned SQUID magnetometer, cooled in a modified liquid helium cryostat. The SQUID input circuit has an intrinsic Q -factor of 28,000, so this is damped by flux-locking electronics and an additional feedback loop to give an effective Q of between 40 and 200 for imaging. The resonant frequency is adjusted by a control rod coupled to a trimmer capacitor mounted in the liquid helium volume. Images of a test object acquired using the cooled receiver exhibit up to two-fold SNR gains in regions close to the sensor, compared to an equivalent room temperature coil, with the noise level dominated by losses coupled from the magnet pole faces. © 2007 IEEE.

Original publication

DOI

10.1109/TASC.2007.898258

Type

Conference paper

Publication Date

01/06/2007

Volume

17

Pages